Forensic Laboratories - Engineering Control Methods
In forensic labs, there is a high concentration of airborne hazards. Biological contamination needs to be on high alert because of the risk associated. Some of the airborne hazards associated include mold, fungus, bacteria, and viruses.
Controlling the spread of these contaminants is crucial in protecting workers from a large variety of illnesses, and potential fatalities in the workplace.
See below for a complete list of engineering control methods for forensic labs.
Controlling the spread of these contaminants is crucial in protecting workers from a large variety of illnesses, and potential fatalities in the workplace.
See below for a complete list of engineering control methods for forensic labs.
For further support, please contact us directly at 1-888-862-5356.
Back & Side Draft Systems
Centrally Ducted Filter Units
Down Draft Tables
Ducted Air Cleaners
Dust Control Booths
Exhaust Fans & Controls
Extraction Arms
Industrial Hose & Duct Work
Lev-Co offers a wide variety of lightweight, flexible, abrasive-resistant, heat-tolerant, innovative technical hoses to provide solutions for a wide range of industrial & automotive applications. Here are some general design velocities when using "CAPTURE" Hoods:
Smoke (e.g. Welding):
- Capture Velocity 100 FPM
- Transport Velocity 3000 FPM
- Minimum Face velocity 2355 FPM
Dust (e.g. Grinding):
- Capture Velocity 500 FPM
- Transport Velocity 4000 FPM
- Minimum Face Velocity 4000 FPM
Gas (e.g. V.O.C.'s):
- Capture Velocity 50 FPM
- Transport Velocity 2000 FPM
- Minimum Face Velocity 785 FPM
Mist:
- Capture Velocity 100 FPM
- Transport Velocity 2250 FPM
- Minimum Face Velocity 1766 FPM
FOR MORE SUPPORT IN SELECTING THE RIGHT HOSE pls EMAIL us with details on your application at: sales@lev-co.com.